VEMCO – VR100 Receiver Operating Manual – Page 13
NOTE: If the animal is operating in deeper water, then adjust the angle of the directional hydrophone so
it’s pointing downward as you get closer to the tag. If you don’t adjust the angle, you might miss
the animal because the hydrophone is looking at the surface of the water instead into its depth.
2.2.6
Enter slope and intercept for a continuous one-sensor transmitter
The tag’s information that must be entered in the VR100 is found in the Transmitter Specification manual
shipped with the tags. See section 4.3.4 for more details.
From the Main menu:
1.
Select Config (right selection button).
2.
Press “4” on the keypad.
3.
Press “2” on the keypad.
4.
If a list of serial numbers appears, press the left selection button to select “New”. Otherwise, go to
Step 5.
5.
Enter the serial number of the tag and press “Next”.
6.
Enter the tag’s operating frequency.
7.
Enter the tag’s slope value.
8.
Enter the tag’s intercept value.
9.
Select the data type, either temperature or depth.
10.
Enter the Minimum Period value.
11.
Enter the Maximum Period value.
12.
Press the right selection button to “Accept” this tag.
13.
If more tags are to be entered, repeat steps 3 to 12.
14.
Press the Menu button until the Main screen appears.
2.2.7
Enter slope and intercept for a continuous two-sensor transmitter
The tag’s information that must be entered in the VR100 is found in the Transmitter Specification manual
shipped with the tags. See section 4.3.4 for more details.
From the Main menu:
1.
Select Config (right selection button).
2.
Press “4” on the keypad.
3.
Press “3” on the keypad.
4.
If a list of serial numbers appears, press the left selection button to select “New”. Otherwise, go to
Step 5.
5.
Enter the serial number of the tag and press “Next”.
6.
Enter the tag’s operating frequency.
7.
Enter the tag’s sync value (1150 is the default value).
8.
Enter the tag’s first slope value.
9.
Enter the tag’s first intercept value.
10.
Select the data type for the first sensor (usually temperature).
11.
Enter the tag’s second slope value.
12.
Enter the tag’s second intercept value.
13.
Select the data type for the second sensor (usually depth).