70
11 Supplement
VEGABAR 82 • Profibus PA
45031-EN-220624
Process temperature - Process fittings plastic
Measuring cell seal
Process temperature
Process fitting PEEK
8)
Process fitting PVDF
9)
FKM
VP2/A
-20 … +100 °C (-4 … +212 °F)
-20 … +80 °C (-4 … +176 °F)
10)
A+P 70.16
-40 … +100 °C (-40 … +212 °F)
EPDM
A+P 70.10-02
FFKM
Kalrez 6375
-20 … +100 °C (-4 … +212 °F)
Perlast G74S
-15 … +100 °C (5 … +212 °F)
Perlast G75B
Temperature derating
1
2
0 °C
(32 °F)
-40 °C
(-104 °F)
55 °C
(131 °F)
80 °C
(176 °F)
130 °C
(266 °F)
100 °C
(212 °F)
-40 °C
(-104 °F)
Fig. 44: Temperature derating VEGABAR 82, version up to +130 °C (+266 °F)
1 Process temperature
2 Ambient temperature
1
2
0 °C
(32 °F)
-40 °C
(-104 °F)
50 °C
(122 °F)
80 °C
(176 °F)
150 °C
(302 °F)
100 °C
(212 °F)
-40 °C
(-104 °F)
Fig. 45: Temperature derating VEGABAR 82, version up to +150 °C (+302 °F)
1 Process temperature
2 Ambient temperature
8)
Max. permissible process pressure: 30 bar
9)
Max. permissible process pressure: 10 bar
10)
Process pressures > 5 bar: 20 … +60 °C (-4 … +140 °F)
Содержание VEGABAR 82
Страница 94: ...94 Notes VEGABAR 82 Profibus PA 45031 EN 220624 ...