9 Diagnosis, asset management and service
VEGABAR 82 • Profibus PA
45031-EN-220624
Code
Text message
Cause
Rectification
DevSpec
Diagnosis Bits
F013
No valid measured value
available
Gauge pressure or low pressure
Measuring cell defective
Exchange measuring cell
Send instrument for repair
Bit 0
F017
Adjustment span too
small
Adjustment not within specifi-
cation
Change the adjustment accord-
ing to the limit values
Bit 1
F025
Error in the lineariza-
tion table
Index markers are not continu-
ously rising, for example illogical
value pairs
Check linearization table
Delete table/Create new
Bit 2
F036
no operable sensor soft-
ware
Failed or interrupted software
update
Repeat software update
Check electronics version
Exchanging the electronics
Send instrument for repair
Bit 3
F040
Error in the electronics
Hardware defect
Exchanging the electronics
Send instrument for repair
Bit 4
F041
Communication error
No connection to the sensor
electronics
Check connection between sen-
sor and main electronics (with
separate version)
Bit 13
F042
Communication error
Secondary sensor
No connection to the Second-
ary sensor
Check connection between Pri-
mary and Secondary sensor
Bit 28 of
Byte 0 … 5
F080
General software error
General software error
Disconnect operating voltage
briefly
Bit 5
F105
Measured value is de-
termined
The instrument is still in the
switch-on phase, the measured
value could not yet be deter-
mined
Wait for the end of the switch-
on phase
Bit 6
F113
Communication error
Error in the internal instrument
communication
Disconnect operating voltage
briefly
Send instrument for repair
Bit 12
F260
Error in the calibration
Error in the calibration carried out
in the factory
Error in the EEPROM
Exchanging the electronics
Send instrument for repair
Bit 8
F261
Error in the instrument
settings
Error during setup
Error when carrying out a reset
Repeat setup
Repeat reset
Bit 9
F264
Installation/Setup error
Inconsistent settings (e.g.: dis-
tance, adjustment units with
application process pressure) for
selected application
Invalid sensor configuration (e.g.:
application electronic differential
pressure with connected differen-
tial pressure measuring cell)
Modify settings
Modify connected sensor config-
uration or application
Bit 10
Failure