![Vega VEGABAR 82 Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/vega/vegabar-82/vegabar-82_operating-instructions-manual_859602013.webp)
13
3 Product description
VEGABAR 82 • Profibus PA
45031-EN-220624
The front-flush, hygienic installation of the measuring cell acc. to 3A
is particularly suitable for food applications. The sealings are installed
gap-free. The additional front sealing for the measuring cell protects
also the glass joint. A hole in the process fitting is used for leakage
detection.
1
6
2
3
4
5
Fig. 10: Hygienic installation of the measuring cell acc. to 3-A (example: Clamp
connection)
1 Measuring cell
2 Seal for the measuring cell
3 Process fitting
4 Additional, front seal for measuring cell
5 Diaphragm
5 Hole for leakage detection
3.3 Supplementary cleaning procedures
The VEGABAR 82 is also available in the version "
Oil, grease and
silicone-free
" or cleaning for paint compatible version (LABS). These
instruments have passed through a special cleaning procedure to
remove oil, grease and paint-wetting impairment substances (PWIS).
The cleaning is carried out on all wetted parts as well as on surfaces
accessible from outside. To keep the purity level, the instruments are
immediately packed in plastic foil after the cleaning process. The
purity level remains as long as the instrument is kept in the closed
original packaging.
Caution:
The VEGABAR 82 in these versions must not be used in oxygen
applications. For this purpose, devices in the special version "
Resistance to Internal Ignition in oxygen operation according to BAM
certification
" are available.
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging consists of environment-friendly, recyclable card-
board. For special versions, PE foam or PE foil is also used. Dispose
of the packaging material via specialised recycling companies.
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con-
cealed defects must be appropriately dealt with.
Installation in hygienic
fitting acc. to 3-A
Packaging
Transport
Transport inspection