11 Diagnostics and servicing
FIBERTRAC 31 • Profibus PA
43833-EN-170126
Code
Text mes-
sage
Reason
Removal
F121
Faulty partici
-
pant list on the
multisensor
communica-
tion bus
•
Slave instruments not found
•
Check Slave instruments
•
Check Slave list in Master
instrument
•
Slave instrument with wrong
address
F122
Double ad
-
dresses on
the mul
-
tisensor
communica-
tion bus
•
Instrument addresse was
assigned several times
•
Change instrument
addresses
F123
X-ray alarm
•
External instruments cause
radiation
•
Determine reason for X-ray
alarm
•
In case of brief X-ray radia
-
tion: Monitor switching out
-
puts for this time manually
F124
Alarm due to
increased ra
-
diation
•
Radiation dose too high
•
Determine reason for
increased radiation
F125
Ambient tem
-
perature too
high
•
Ambient temperature on
the housing outside the
specification
•
Cool the instrument or pro
-
tect it with isolating material
against radiation heat
The following table shows the error codes and text messages in the
status message "
Function check
" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Reason
Removal
C029
Simulation
•
Simulation active
•
Finish simulation
•
Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "
Out of specification
" and provides information on
causes as well as corrective measures.
Code
Text mes-
sage
Reason
Removal
S017
Accuracy
outside the
specification
•
Accuracy outside the speci
-
fication
•
Correct adjustment data
Function check
Out of specification