![Vega FIBERTRAC 31 Скачать руководство пользователя страница 70](http://html2.mh-extra.com/html/vega/fibertrac-31/fibertrac-31_operating-instructions-manual_3808664070.webp)
70
11 Diagnostics and servicing
FIBERTRAC 31 • Profibus PA
43833-EN-170126
still valid. Plan in maintenance for the instrument because a failure is
expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
The following table shows the error codes and text messages in the
status message "
Failure
" and provides information on causes as well
as corrective measures.
Example for a failure message
Code
Text mes-
sage
Reason
Removal
F008
Error multi
sensor com-
munication
•
Additional sensors not
switched on
•
EMC influences
•
No other sensor available
•
Check wiring between the
sensors
•
Connect the sensors cor-
rectly and make them ready
for operation
F013
Sensor sig-
nals a fault
•
Error on the current input
•
No valid measured value
•
Connected instruments
without function
•
Check current input
•
Check connected instru
-
ments (Slaves)
F016
Adjustment
data ex
-
changed
•
Values of the min. and max.
adjustment exchanged
•
Correct adjustment data
F017
Adjustment
span too small
•
The values of the min. and
max. adjustment are too
close together
•
Correct adjustment data
F025
Invalid lineari
-
zation table
•
Empty linearization table
•
Wrong value in the lineariza
-
tion table
•
Create linearization table
•
Correct linearization table
F030
Process value
out of limits
•
Process values are not
within the adjusted measur
-
ing range
•
Repeat adjustment
F034
EPROM hard
-
ware error
•
Electronics defective
•
Restart instrument
•
Exchanging the electronics
F035
EPROM data
error
•
Error in the internal instru
-
ment communication
•
Carry out a reset
•
Exchanging the electronics
F036
Faulty pro
-
gram memory
•
Error during software update
•
Repeat software update
•
Exchanging the electronics