6 Set up with the display and adjustment module
FIBERTRAC 31 • Profibus PA
43833-EN-170126
For this purpose, check which isotope is in the source container. You
can find this information on the type label of the source container.
With this selection, the sensitivity of the sensor is adapted in an opti
-
mum way to the isotope.
The FIBERTRAC 31 requires this information for the decay compen
-
sation. An annual recalibration is hence not necessary.
Enter the requested parameters via the appropriate keys, save your
settings with
[OK]
and jump to the next menu item with the
[ESC]
and
the
[->]
key.
Enter here, the respective application.
This menu item enables adaptation of the sensor to the requested ap
-
plication. You can choose between the following applications: "
Level
",
"
Point level
" or "
Summation Slave
".
In this menu item you can activate the function of the current output.
When the output is activated, the instrument remains in its function as
a Slave, but the 4 … 20 mA output of the FIBERTRAC 31 can be also
used als single instrument.
When the output is active, the instrument has the complete function
-
ality of a level measuring instrument. In this case, continue reading
under "
Parameter adjustment/Level measurement
".
Since the adjustment is very comprehensive, the menu points of
Function Blocks 1 (FB1) were put together in a submenu.
In menu item"
Channel
" you determine which measured value the
output refers to.
Setup - Application
Setup - Outputs
Setup - AI FB1
Setup - AI FB1 - Channel