Notice to the Installers
Installation Site
1. PC monitors, emergency shutdown switches, and X-ray exposure switches
should be installed near the operator for emergency management.
2. Proper shielding of the room is essential: Since these requirements vary
depending on the country, it is the installer’s responsibility to verify that all
applicable radiation safety requirements are met.
3. This equipment should not be installed in the immediate vicinity of other
devices.
4. Do not install the equipment in an area that is exposed to strong
electromagnetic fields.
5. Do not install this system in an area where there is the risk of an explosion.
6. The electrical installation of this system shall comply with all local code
requirements for electro-medical systems:
IEC 60364-7-710
.
7. It is strongly recommended that a UPS be installed at the same time as the
equipment.
8. The equipment, PC, and all peripheral devices must be well-grounded.
Warnings Regarding X-Ray Radiation
1. Failure to install this equipment in an approved location may be dangerous
to the patient and operator.
2. Stationary radiation shielding must be installed to protect the operator from
radiation.
3. The X-Ray system may cause injury to the patient if improperly used. Obey
all federal and municipal standards regarding radiation safety.
4. When exposing the patient to the X-Ray, the operator must be behind a
protective wall or take other protective actions. The operator should remain
at least 2 m (7 feet) away from the X-Ray when pressing the exposure
switch and observe the patient and capture-progression.
5. Operators must provide protective clothing to the patient before X-Ray
capturing. Pregnant women must consult with a doctor prior to being
exposed to an X-Ray.
Содержание Pax-i PCH-2500
Страница 1: ...Installation Manual Full version Model PCH 2500 Version 3 06 English...
Страница 2: ......
Страница 3: ......
Страница 46: ...This page is intentionally left blank...
Страница 82: ...This page is intentionally left blank...
Страница 104: ...This page is intentionally left blank...
Страница 124: ...This page is intentionally left blank...
Страница 166: ...This page is intentionally left blank...
Страница 167: ...Acquiring the Test Image 10...
Страница 169: ...11 Technical Specifications...
Страница 176: ...This page is intentionally left blank...
Страница 179: ...PaX i PCH 2500 Installation Manual 179 Appendix Schematic diagram Components supplied...
Страница 183: ...PaX i PCH 2500 Installation Manual 183 Appendix 2 3M 2 2M 2 1M 2 0M 4 Separate the side cover...
Страница 199: ......
Страница 200: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...