2 Choosing an Installation Site
PaX-i and PC system
No. 12 AWG or larger
Live
Neutral
AC Outlet
Central distribution panel
w/a circuit breaker
Protective Ground (PE)
1. To assure line voltage quality, a separate 3-core grounded power cable connected directly to the
central distribution panel with an over-current circuit breaker rated for 20/15A must be used.
2. The mains resistance should not exceed 0.5 Ω.
3. This equipment should be connected to the earthed outlet.