10 Acquiring the test image
Acquiring the test image
1.
Perform the test image acquisition after the software is installed.
2.
Ensure that the collimator is well aligned.
When the collimator is misaligned, the correct test image can’t be obtained,
in∙which∙case∙the∙alignment∙correction∙must∙be∙performed∙first,∙according∙to∙
the technical manual.
3.
Acquire the test image. For further details about the image acquisition, refer to the accompanying
user manual.
For the equipment with the Cephalometric unit, if the noisy image on the
lower part appears, this is caused by the improper leveling between the CEPH
and the column units. Level the CEPH unit again.
For other issues related to the image, refer to the section(s) regarding X-Ray
alignment in the accompanying service manual.
Assembling the vertical frame cover
Truss bolts
M4 x 8
Part No.24
Qty: 9
Philips screwdriver w/
magnetic tip
L=200 mm(7.9”)
1.
Assemble the vertical top cover and fix it with
7 truss bolts
(Part No.24).
Содержание Pax-i PCH-2500
Страница 1: ...Installation Manual Full version Model PCH 2500 Version 3 06 English...
Страница 2: ......
Страница 3: ......
Страница 46: ...This page is intentionally left blank...
Страница 82: ...This page is intentionally left blank...
Страница 104: ...This page is intentionally left blank...
Страница 124: ...This page is intentionally left blank...
Страница 166: ...This page is intentionally left blank...
Страница 167: ...Acquiring the Test Image 10...
Страница 169: ...11 Technical Specifications...
Страница 176: ...This page is intentionally left blank...
Страница 179: ...PaX i PCH 2500 Installation Manual 179 Appendix Schematic diagram Components supplied...
Страница 183: ...PaX i PCH 2500 Installation Manual 183 Appendix 2 3M 2 2M 2 1M 2 0M 4 Separate the side cover...
Страница 199: ......
Страница 200: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...