PaX-i
(PCH-2500) Installation Manual
79
4 Installing the Equipment: Floor Standing (Optional)
4.7 Leveling the Equipment
Ensure that the spirit level should rest only on the locations indicated in
the∙following∙figures∙to∙obtain∙the∙accurate∙center.
1.
Prepare the spirit level.
X-ray tube head
Rotating
Unit
2.
Turn the Rotating Unit manually until the
X-ray tube head points forward as shown in
the figure.
3.
Turn the Level Foot on the Base Unit
clockwise with the wrench until the equipment
touches the ground as shown in the
illustration.
Allen
wrench