4. Installation Procedures
Green21 (Model: PCT-90LH) Installation Manual
45
EN
G
LISH
* If CSET_1 value is greater than CRDL value, input CRDL value as CSET_1
value.
(ex: CSET_1 value is 43 and CRDL value is 39, enter [SPM_CRDL_43].)
5) Up limit setting
i.
Place the Column at the topmost (using the LCD button.)
ii.
Column Topmost limit setting: [SPM_CLUL]
iii.
Place the Chinrest at the Topmost (using the LCD button.)
iv.
Chinrest Topmost limit setting: [SPM_CRUL]
v.
Rotator 90-degree position movement: [SPM_ARCT] →
[SPM_ROMV_0900]
vi.
Capture CTB and adjust the column height so that the Y value is in the
range of 1 to 3 pixels (0.3 mm to 1 mm) in the projection.
vii.
Topmost Automatic Positioning: [SPM_CSET_2]
6) [SPM_CLEN_1] Input (anti-collision function on)
Содержание Green 21 PCT-90LH
Страница 1: ...User Manual Model PCT 90LH Version 1 49 Full version English...
Страница 2: ......
Страница 13: ...Table of Contents Green21 Model PCT 90LH Installation Manual 1 ENGLISH This Page Left Blank Intentionally...
Страница 14: ...0 Table of Contents 2 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 32: ...3 Before the Installation 20 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 68: ...4 Installation Procedures 56 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 111: ...8 Acquiring a Test Image Green21 Model PCT 90LH Installation Manual 99 ENGLISH This Page Left Blank Intentionally...
Страница 115: ...9 Technical Specifications Green21 Model PCT 90LH Installation Manual 103 ENGLISH This Page Left Blank Intentionally...
Страница 127: ......
Страница 128: ......