7. Installing Software
Green21 (Model: PCT-90LH) Installation Manual
93
EN
G
LISH
Disabling the packing mode
Green21 has a packing mode built into the system to prevent
the unit from being damaged while shipping and transporting.
Thus, it is in the packing mode by factory default. Disabling
the packing mode is required at this step for successful
installation.
Unless the packing mode is disabled, no operation will
happen even after the equipment is turned on.
1. Click on the
General
tab in the control panel.
2. In the
Networking option
, select the
Serial
checkbox and set
Serial port
and
Baud Rate
as follows and then click on
Manager
.
If an error has occurred during connection, make sure the
COM port setting is correct as follows:
1) Run Device Manager.
2) Check the COM port assigned to CONTACT-US Virtual
Serial Port under Ports (COM & LPT).
3) If necessary, change the COM port in the
Networking
option
and click on
Save
.
3. Enter the command
PVER]
to verify the current mode. Note that the equipment is
now in the packing mode.
Figure 7-35. Checking the packing mode
Содержание Green 21 PCT-90LH
Страница 1: ...User Manual Model PCT 90LH Version 1 49 Full version English...
Страница 2: ......
Страница 13: ...Table of Contents Green21 Model PCT 90LH Installation Manual 1 ENGLISH This Page Left Blank Intentionally...
Страница 14: ...0 Table of Contents 2 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 32: ...3 Before the Installation 20 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 68: ...4 Installation Procedures 56 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 111: ...8 Acquiring a Test Image Green21 Model PCT 90LH Installation Manual 99 ENGLISH This Page Left Blank Intentionally...
Страница 115: ...9 Technical Specifications Green21 Model PCT 90LH Installation Manual 103 ENGLISH This Page Left Blank Intentionally...
Страница 127: ......
Страница 128: ......