4. Installation Procedures
Green21 (Model: PCT-90LH) Installation Manual
53
EN
G
LISH
8. Ceph (Carpus) - 4% cut align work
1) Perform 4% cut according to the following criteria.
Pano: Threshold – 80
Collimator
Min
Max
Left (T-80%)
3 pixels
5 pixels
Right (T-80%)
2 pixels
4 pixels
Top (T-80%)
275 pixels 290 pixels
Bottom (T-80%)
15 pixels
30 pixels
2)
Run C:\VCaptureSW\Util\CollimatorAlign\CollimatorAlign.exe file.
Collimator value should be saved after proceeding to Capture
after clicking the Initial button, so Should shoot more than
once for each FOV.
9. Collimator setting value by mode
1) Modes that use the same Collimator value with Left, Right, and Bottom
CT & Ceph
- Dental CT: 21x19, 17x15, 12x9, 8x8
- ENT CT: 21x19, 17x11
- Ceph: LAT, PA, SMV, WAT
Pano & Ceph(Carpus)
- Pano(Adult), Pano(Child), Ceph(CAR)
Содержание Green 21 PCT-90LH
Страница 1: ...User Manual Model PCT 90LH Version 1 49 Full version English...
Страница 2: ......
Страница 13: ...Table of Contents Green21 Model PCT 90LH Installation Manual 1 ENGLISH This Page Left Blank Intentionally...
Страница 14: ...0 Table of Contents 2 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 32: ...3 Before the Installation 20 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 68: ...4 Installation Procedures 56 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 111: ...8 Acquiring a Test Image Green21 Model PCT 90LH Installation Manual 99 ENGLISH This Page Left Blank Intentionally...
Страница 115: ...9 Technical Specifications Green21 Model PCT 90LH Installation Manual 103 ENGLISH This Page Left Blank Intentionally...
Страница 127: ......
Страница 128: ......