2. Choosing an Installation Site
Green21 (Model: PCT-90LH) Installation Manual
7
EN
G
LISH
2. Choosing an Installation Site
2.1
Room Requirements
1. The location of this equipment should allow for high
visibility of the patient by the operator and the operator
should be as about the patient as possible.
2. This equipment should not be installed on thick carpets
for stability reasons.
3. Anti-static floor materials should be used around the
equipment.
4. The PC monitor, emergency cut off switch and X-ray
exposure switch should be installed near the operator so
that he or she can manage them immediately in case of
emergency.
1. Minimum space required:
1527mm/60”
1017mm/40” (Open)
593mm/23”
510mm/20”
1575m
m/
62”
(O
pe
n)
1365m
m/
54
”
856mm
/34
”
2500m
m/
98
”
2000mm/79”
Содержание Green 21 PCT-90LH
Страница 1: ...User Manual Model PCT 90LH Version 1 49 Full version English...
Страница 2: ......
Страница 13: ...Table of Contents Green21 Model PCT 90LH Installation Manual 1 ENGLISH This Page Left Blank Intentionally...
Страница 14: ...0 Table of Contents 2 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 32: ...3 Before the Installation 20 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 68: ...4 Installation Procedures 56 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 111: ...8 Acquiring a Test Image Green21 Model PCT 90LH Installation Manual 99 ENGLISH This Page Left Blank Intentionally...
Страница 115: ...9 Technical Specifications Green21 Model PCT 90LH Installation Manual 103 ENGLISH This Page Left Blank Intentionally...
Страница 127: ......
Страница 128: ......