4.10 Check X-ray align (4% Cut)
Make sure to remove the inner Carbon Cover of the Rotating
Unit before conducting this section.
* Disassembly of the inner Carbon Cover.
Make sure the X-ray align is correct. The criteria for left, right, top and bottom are
as follows.
1) CT: Threshold - 95
Collimator
Min
Max
Left (T-80%)
6 pixels
8 pixels
Right (T-80%)
4 pixels
6 pixels
Top (T-80%)
4 pixels
10 pixels
Bottom (T-80%)
4 pixels
10 pixels
2) Pano: Threshold - 80
Collimator
Min
Max
Left (T-80%)
3 pixels
5 pixels
Right (T-80%)
2 pixels
4 pixels
Top (T-80%)
275 pixels 290 pixels
Bottom (T-80%)
15 pixels
30 pixels
Содержание Green 21 PCT-90LH
Страница 1: ...User Manual Model PCT 90LH Version 1 49 Full version English...
Страница 2: ......
Страница 13: ...Table of Contents Green21 Model PCT 90LH Installation Manual 1 ENGLISH This Page Left Blank Intentionally...
Страница 14: ...0 Table of Contents 2 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 32: ...3 Before the Installation 20 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 68: ...4 Installation Procedures 56 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 111: ...8 Acquiring a Test Image Green21 Model PCT 90LH Installation Manual 99 ENGLISH This Page Left Blank Intentionally...
Страница 115: ...9 Technical Specifications Green21 Model PCT 90LH Installation Manual 103 ENGLISH This Page Left Blank Intentionally...
Страница 127: ......
Страница 128: ......