9. Technical Specifications
Green21 (Model: PCT-90LH) Installation Manual
101
EN
G
LISH
9.2
Electrical Specifications
Item
Description
Power supply voltage
AC 100 ~ 240 V
Frequency
50 / 60 Hz
Power rating
1.3 - 1.6 kVA
The input line voltage depends on the local electrical distribution system.
Allowable input voltage fluctuation requirement: ± 10 %
Mode of operation: non-continuous operation (NFPA 70: long time operation) -
needs waiting time (at least 60 times the exposure time) before the next exposure
begins
Exposure time: Max. 18 s / Resting time: 10 min.
Column operation time: 1 min. / Resting time: 9 min.
To assure line voltage quality, a separate 3-core
grounded power cable connected directly to central
distribution panel with an over-current circuit breaker
rated for 20/15A must be used.
The mains resistance should not exceed 100 V at 0.045-
ohm, 240 V at 0.19 ohm.
Содержание Green 21 PCT-90LH
Страница 1: ...User Manual Model PCT 90LH Version 1 49 Full version English...
Страница 2: ......
Страница 13: ...Table of Contents Green21 Model PCT 90LH Installation Manual 1 ENGLISH This Page Left Blank Intentionally...
Страница 14: ...0 Table of Contents 2 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 32: ...3 Before the Installation 20 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 68: ...4 Installation Procedures 56 Green21 Model PCT 90LH Installation Manual This Page Left Blank Intentionally...
Страница 111: ...8 Acquiring a Test Image Green21 Model PCT 90LH Installation Manual 99 ENGLISH This Page Left Blank Intentionally...
Страница 115: ...9 Technical Specifications Green21 Model PCT 90LH Installation Manual 103 ENGLISH This Page Left Blank Intentionally...
Страница 127: ......
Страница 128: ......