Chapter 6 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 81
To calibrate, do the following:
1.
Make sure that the window is cleaned.
2.
Then open the command line and enter the Advanced level by
giving the
advanced
command.
3.
Give the
set factory win_clean
command.
The calibration is carried out automatically.
NOTE
If the window is visibly worn out, replace it.
Checking the Door Gasket
The door of the measurement unit utilizes an electrically conductive
rubber gasket to suppress electromagnetic radiation. When the door is
opened, check that the gasket and the opposite contact surface is clean.
Use a wet cloth for cleaning if necessary.
Battery Check
Check the battery condition annually. If any signs of aging such as a
bulging battery case, white powder or residue near the battery vent,
leaking electrolyte, or corroded terminals is observed replace the
battery. Lead acid batteries may age in 3 to 5 years time and result in a
rupture and loss of electrolyte.
NOTE
When disposing of old batteries, be sure to follow local
environmental regulations.
NOTE
In freezing temperatures there is a danger of battery rupture if the
battery is completely discharged. Do not store empty batteries in
freezing temperatures. Replace the battery if signs of mechanical
rupture are observed.