Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 43
Table 6
Advanced Level Commands
Command
Description
back
Go back one security level. Go back to the user level
commands.
get failure diag
Show failure diagnostics data.
get params algorithm
Prints cloud algorithm related parameters.
service replace_instr battery
Prints service instructions for replacing a battery.
service replace_instr
cle_engine_board
Prints service instructions for replacing a CLE311
engine board.
service replace_instr clp_ac_power
Prints service instructions for replacing a CLP311
power unit.
service replace_instr clr_receiver
Prints service instructions for replacing a CLR311
receiver.
service replace_instr clt_transmitter
Prints service instructions for replacing a CLT311
transmitter.
service self_check
Runs the self-check.
service spare_part cle_engine mark
Marks CLE spare part status.
service spare_part cle_engine clear
Clears CLE spare part status.
service spare_part clt_transmitter
mark
Marks CLT spare part status.
service spare_part clt_transmitter
clear
Clears CLT spare part status.
service sw_update
Updates the software.
set data_acq autoadj on
Sets SW-control of data acquisition parameters.
set data_acq autoadj off
Disables SW-control of data acquisition parameters.
set data_acq meas_mode standard
Maximum range 7700 m, 10 m resolution, laser
pulse rate 10 kHz.
set data_acq meas_mode high_res
Maximum range 7550 m, 5 m resolution, laser pulse
rate 8 kHz.
set data_acq power_save disable
Disables power-save mode.
set data_acq power_save enable
Enables power-save mode.
set data_acq power_save interval
30 ... 30000
Power-save interval in seconds. Default is 60.
set data_acq receiver gain low
Sets receiver to low gain.
set data_acq receiver gain high
Sets receiver to high gain.
set data_acq transmit length_of_p short Sets transmitter to short pulse.
set data_acq transmit length_of_p long Sets transmitter to long pulse.
set data_acq transmit inlaser 0 ...
4095
Sets control value for laser pulse.
set factory outlaser 0 .. 2500
Sets target outlaser.
set factory win_clean
Sets window clean status for window cond.
calibration.
set message angle_corr on
Heights in messages are corrected for the tilt angle.
set message angle_corr off
Heights in messages are not corrected for the tilt
angle.
set message height_offset <value>
Sets height offset in current units.
(Range is -304 ... 304 m or -1000 ... 1000 ft.)
Positive values are added to and negative values are
subtracted from the measured height.
set message manual_msg <string>
Sets a manual message. Empty string disables.
set message profile scale
Scaling factor for range gate data (%).
set message profile noise_h2 on
Range gates data is always range normalized, even
noise.