Chapter 8 ___________________________________________________________________ Repair
VAISALA ______________________________________________________________________ 109
Replacing Internal Cable Set
CAUTION
Servicing the equipment must only be performed by qualified
maintenance personnel.
NOTE
The internal cable set is recommended to be replaced indoors.
To replace the internal cable set, you will need the following tools:
- 2.5 mm Allen-key
- Flat-bladed screwdriver, 3 mm wide tip
- Phillips PH1 screwdriver
Then proceed as follows:
1.
Open the measurement unit door and switch the power off with
all three switches (F1, F2, and battery). For the location of the
switches, refer to Figure 18 on page 98 and Figure 19 on page
99.
WARNING
Disconnect the CT3839 power cable before continuing!
2.
Close the door and remove the measurement unit. To do this,
loosen the three attachment screws (marked A in Figure 5 on
page 23) on the back of the measurement unit, disconnect the
blower cable from connector J1 (see Figure 7 on page 25), and
pull out the unit.
3.
The internal cable set is attached to the back wall of the
measurement unit enclosure, thus to gain access to it you have to
remove the CLE311 board, the optics assembly with the
transmitter, and the receiver. Refer to sections Replacing
Ceilometer Engine Board CLE311 on page 100, and Replacing
Ceilometer Optics CLO311 on page 105. Remove the connector
of the cable set from the CLP311 unit.
4.
Remove the grounding screw from the back of the optics frame
by loosening it and thus disconnecting the two protective
grounding terminals attached to it. See Figure 20 on page 110.