Chapter 8 ___________________________________________________________________ Repair
VAISALA ______________________________________________________________________ 105
8.
Switch the power on with all three switches. Wait for 3 minutes
after the start-up until self-check is completed. Ensure that all
six diagnostic LEDs are lit. Refer to Figure 19 on page 99.
The ceilometer unit should now return to normal operation. Verify
that the unit is working properly and that there are no other failures
with the system. If other failures exist, separate troubleshooting may
be necessary.
Replacing Ceilometer Optics CLO311
CAUTION
Servicing the equipment must only be performed by qualified
maintenance personnel.
NOTE
Ceilometer Optics CLO311 is recommended to be replaced indoors.
To replace Ceilometer Optics CLO311, you will need a 3 mm Allen-
key.
Then proceed as follows:
1.
Open the measurement unit door and switch the power off with
all three switches (F1, F2, and Battery). For the location of the
switches, refer to Figure 18 on page 98 and Figure 19 on page
99.
WARNING
Disconnect the CT3839 power cable before continuing!
2.
Close the door and remove the measurement unit. To do this,
loosen the four attachment screws (marked A in Figure 5 on
page 23) on the back of the measurement unit, disconnect the
blower cable from connector J1 (see Figure 7 on page 25), and
pull out the unit.
3.
Disconnect the battery, transmitter, receiver, laser monitor
board, and data cables from the CLE311 board.
4.
To remove the Ceilometer Optics assembly, first remove the
optics holder by loosening the four Allen screws on its front