Chapter 9 ____________________________________________________________ Technical Data
VAISALA ______________________________________________________________________ 119
External Connector J3 - Data Line
The data line is intended to be used for measurement data
communication, but it can also be used with a Ceilometer
Maintenance Terminal, PC, or other terminals.
Table 24
Data Line
Property
Description / Value
Connector (J3)
Type Binder series 693, 09-4227-00-07
7-pin (male)
Mating connector type
Type Binder series 693, 99-4226-70-07
(female) elbow
Baud rate
9600 bps standard with RS-232
300, 600, 1200, 2400, 4800, 9600, 19200,
38400, 57600 bps available
300, 1200 or 2400 bps available with
modem DMX501
Max. distance to operate
300m (1000 ft) with RS-232,
1.2 km (4000 ft) with RS-485, all at 2400
bps with typical communication cables
16 km (10 mi) with Modem
Standard character frame
1 Start Bit
8 Data Bits
No Parity
1 Stop Bit
Standard character code
USASCII
Pin connections
1 RS485 -/RTS# (no SW support)
2 RS485+/RS-232 RxD/ Modem Lead A
3 Modem Lead B /RS-232 TxD
4 Signal Ground
5 CTS# (no SW support)
6 +12 V DC (700 mA continuous, for local
external equipment)
Modem circuits are non-polar and symmetrical. All modem circuits
are electrically floating to overvoltage protection rating
(300 V - 500 V).
Overvoltage protection in
each circuit
Primary Noble Gas Surge Arrester
Secondary VDRs, Transient Zener Diodes
or normal Diodes
NOTE