Documents are only to be used and distributed completely and unchanged, especially including the section ”Safety information”. This is page 4 of a document consisting of
49 pages. The instructions for use describe only the actual product. If the product is older or newer than the date of writing, it is strictly the users´ responsibility to check carefully
the validity of the document (if in doubt, order appropriate instructions for use). 99 90 34 / 12/09/2003
Safety information!
Remove all packing material, remove the product from its packing-box, remove the
protective covers from the inlet and outlet ports and keep, inspect the equipment.
If the equipment is damaged, notify the supplier and the carrier in writing within three
days; state the item number of the product together with the order number and the
supplier’s invoice number. Retain all packing material for inspection.
Do not use the equipment if it is damaged.
If the equipment is not used immediately, replace the protective covers. Store the
equipment in suitable conditions.
☞
Read and obey this manual before installing or operating the equipment.
☞
Transport the pump at the provided handles.
Use the equipment
for the intended use only
(for generation and measurement of
vacuum).
☞
☞
☞
☞
☞
Prevent any part of the human body from coming in contact with the vacuum.
☞
Obey notes on correct vacuum and electrical connections, see section ”Use and
operation”.
☞
Make sure that the individual components are only connected, combined and op-
erated according to their design and as indicated in the instructions for use.
Obey
national safety regulations and safety requirements
concerning the use of
vacuum and electrical equipment.
☞
The mains switch for the controller and the pump is at the rear side of the control-
ler.
☞
After switching off the pump, wait 60 sec. minimum until next switching on.
☞
Due to the
high leakage current (> 3.5 mA)
, a too low rated frequency converter
safety switch may be activated, especially if several pumps are operated parallelly.
Attention: Make sure that a protective conductor connection is established be-
fore the equipment is connected to mains supply!
➨
The pumping units MZ 2C VARIO, MD 4C VARIO, PC 2002 VARIO, PC 2003 VARIO
und PC 2004 VARIO are devices with double-pole cutoff and a symmetric capaci-
tive circuit.
☞
If the leakage current is obtained by measuring the equivalent leakage current (ac-
cording to VDE 0701, September 2000, section 5.7), the measured value may be
devided by a factor of 2 in case the device is equipped with a double-pole cutoff and
a symmetric capacitive circuit.
☞
The shock currents according to EN 61010 are below the limit of 3,5 mA.
☞
Equipment must be connected only to a
suitable fused and protected electrical
supply
and a suitable earth point. Failure to connect the motor to ground may
result in deadly electrical shock.
☞
The supply cable may be fitted with a moulded European IEC plug or a plug suit-
able for your local electrical supply. If the plug has been removed or has to be
removed, the cable will contain wires colour coded as follows: green or green and
yellow: earth; blue or white: neutral; brown or black: live.
☞
Check that mains voltage and current conform with the equipment (see rating plate).
☞
If the equipment is brought from cold environment into a room for operation, allow
the equipment to warm up (pay attention to water condensation on cold surfaces).
☞
Make sure ventilation is adequate if pump is installed in a housing or if ambient
temperature is elevated.
Obey all
relevant safety requirements
(regulations and guidelines) and adopt suitable
safety measures.