Documents are only to be used and distributed completely and unchanged, especially including the section ”Safety information”. This is page 21 of a document consisting of
49 pages. The instructions for use describe only the actual product. If the product is older or newer than the date of writing, it is strictly the users´ responsibility to check carefully
the validity of the document (if in doubt, order appropriate instructions for use). 99 90 34 / 12/09/2003
If the automatic mode has been preselected
“End”
is displayed.
☞
Switching between automatic switching off activated or not
activated is possible by pressing key
▲
or
▼
.
➨
Confirm by pressing key START/STOP.
☞
Automatic switching off as soon as the controller detected
that evaporation is completely finished. The pump is switched
off, the cooling water is running on, if a cooling water valve
is preselected.
☞
Without automatic switching off, the pressure in the vacuum
system is adjusted after evaporation is completed if neces-
sary until the ultimate vacuum of the system.
The controller switches to normal operation mode (process control not active).