Interface parameters
The controller CVC 2000
II
is equipped with a serial interface at the rear side of the housing (RS 232C, nine-
pole Sub-D-plug).
☞
Respectively plug-into or remove the cable (null modem cable) from the interface only if the equipment
is switched off.
☞
The interface is not electrically isolated from the measuring circuit.
☞
For optimal electromagnetic compatibility assemble an interface filter (cat. no.: 63 82 35).
The controller can be operated via serial interface. Measuring results, preselections and the status of the
controller can be read at any time. Controlling via interface is only possible, if the remote operation mode
has been preselected at the controller.
During remote operation (”PC” is displayed) the controller
can be operated only via interface, the keys at the controller have no function, the warning triangle
is displayed.
Setting of the interface
Setting of the interface parameters directly at the controller is described below. The factory set values are
underlined.
➨
1200, 2400, 4800 or 9600 BAUD
➨
7 data bits odd, 7 data bits even, 8 data bits none, 8 data bits odd, 8 data bits even
➨
no Handshake, XON/XOFF Handshake, CTS/RTS Handshake
➨
remote on, remote off
☞
Unchangeable: Startbit = 1, Stopbit = 1
Press key Mode
▲
during switching on.
☞
The controller switches to the mode to configurate the inter-
face.
BAUD-Rate is displayed.
➨
Press key Mode
▲
or
▼
to select the BAUD rate and con-
firm with key STOP (”96”=9600, ”48”=4800, ”24”=2400,
”12”=1200).
Data bits are displayed.
➨
Press key Mode
▲
or
▼
to select the data bits and confirm
with key STOP.
Parity is displayed.
➨
Press key Mode
▲
or
▼
to select parity and confirm with
key STOP (”no” = no check of parity, ”O” = odd, ”E” = even).
Handshake is displayed.
➨
Press key Mode
▲
or
▼
to select handshake and confirm
with key STOP (”no” = no Handshake, ”HA” = RTS/CTS,
”SO” = XON/XOFF).
Remote is displayed.
➨
Press key Mode
▲
or
▼
to select parity and confirm with
key STOP (”no” = local operation, ”On” = remote operation).
☞
The controller switches to normal operation mode (process
control not active).
+
(mains switch)
Set
b.96
Set
d. 8
Set
P.no
Set
h.no
Set
r.no