Assembling the pump at the pump support on the base plate
Position pump with pump support on the feet of the pump sup-
port and screw on four cap nuts.
Montage of the connecting hose
Individual performance check of a pump head
By measuring the pressure at the inlet port of the individual
head:
Use a suitable vacuum gauge (e. g. DVR 2, cat. no.: 68 29 02),
make sure that it is correctly calibrated, and measure the pres-
sure at the inlet port. A vacuum of less than 90 mbar should be
indicated.
☞
If the reading is higher, recheck the pump chamber and make
sure that the valves and the diaphragms are correctly seated
(diaphragms concentric with bore).
Use open ended wrench (w/f 15) to reconnect hose to elbow
fitting.
Tighten union nuts first by hand and then tighten one full turn
using open ended wrench (w/f 17).
Reassemble in reverse order.
Install head cover, valves and housing cover insert with hous-
ing cover.
☞
Make sure that the valves are correctly seated.
Screw in four socket head screws fixing housing cover cross-
wise (e. g. in the sequence
➀
,
➁
,
➂
,
➃
) first slightly, then tighten.
☞
Do not tighten until head cover is in contact with housing,
max. torque 10 Nm.
➂
➂
➂
➂
➂
➃
➃
➃
➃
➃
➀
➀
➀
➀
➀
➁
➁
➁
➁
➁