Tighten union nut first by hand and then tighten by one full turn
using open-ended wrench.
Reconnect hose to the inlet of the condenser and tighten union
nut.
Use open-ended wrench to reconnect hose to elbow fitting.
If the pump does not achieve the ultimate pressure:
☞
In case the diaphragms and valves have been replaced, a run-in period of several
hours is required before the pump achieves its ultimate vacuum.
If all pump heads achieve a vacuum below 90 mbar but pump does not achieve the
ultimate total pressure:
Check hose connectors between pump heads and manifolds for leaks. If necessary
recheck pump chamber.