Documents are only to be used and distributed completely and unchanged, especially including the section ”Safety information”. This is page 20 of a document consisting of
49 pages. The instructions for use describe only the actual product. If the product is older or newer than the date of writing, it is strictly the users´ responsibility to check carefully
the validity of the document (if in doubt, order appropriate instructions for use). 99 90 34 / 12/09/2003
VACUU•LAN:
☞
Operation of the pump depending on actual pressure de-
mand.
☞
The cooling water is switched off if no more pumping is re-
quired, switched on again if gas or vapour occur.
☞
The mode for decentralized vacuum network systems (for
vacuum in laboratories, vacuum lines for weekend, night or
trainee operations). Continuous vacuum without continuous
pumping.
☞
Additional extension of the lifetime of diaphragms and valves
(only as many diaphragm strokes as necessary).
Setting the mode
Starting the program:
➨
Press key MODE while switching on the controller.
☞
The mode “pressure control” (factory-set) is displayed on
the LCD by flashing symbols.
To change the mode:
➨
Press key
▲
or
▼
until the symbol of the desired mode is
flashing.
Only in mode pressure control:
☞
“Set“, “Auto“ and ”yes” or ”no” are displayed.
➨
Switching between ”yes” or ”no” is possible by pressing key
▲
or
▼
.
➨
To preselect the automatic mode confirm ”yes” by pressing
key START/STOP.
➨
If you do not wish to preselect the automatic mode confirm
”no” by pressing key START/STOP.
pressure control
continuous pumping VACUU•LAN
☞
The symbol of the air admittance valve is displayed and
”yes” or ”no”.
➨
Switching between ”yes” or ”no” is possible by pressing key
▲
or
▼
.
➨
Confirm the selection ”yes” or ”no” by pressing key START/
STOP.
After selecting the mode:
☞
The symbol of the
cooling water valve
is displayed and
”yes” or ”no”.
➨
Switching between ”yes” or ”no” is possible by pressing key
▲
or
▼
.
➨
Confirm the selection ”yes” or ”no” by pressing key START/
STOP.
MODE
+
(mains switch)
➨
Press key START/STOP to confirm the selected mode.