UTD2000M
User
Manual
Table
2
‐
14
Sampling
Menu
Menu
Setting
Description
Acquiring
Mode
Normal
Sampling
Peak
Detection
Average
High
Resolution
Normal
sampling
mode
Peak
detection
mode,
which
is
used
to
detect
burrs
and
to
reduce
the
possibility
of
having
false
waveform;
Set
average
sampling
mode
and
display
average
times;
Set
high
resolution
display;
Sampling
Mode
Equivalent
Real
time
It
requires
multiple
times
of
repeated
sampling
before
acquiring
the
sampling
data
needed
to
form
a
waveform;
Acquiring
all
data
needed
to
form
a
waveform
during
a
single
triggering.
Memory
Depth
Normal
Deep
Memory
Normal
memory
depth:
6kpts
Memory
Depth
Bandwidth
≤
100MHz
16Mpts
or
32Mpts
Bandwidth>100MHz
8Mpts
or
16Mpts
快速采集
Fast
Sampling
On
Off
Fast
sampling
is
enabled
if
it
is
on.
By
fast
sampling,
it
only
takes
samples
with
minimal
memory
depth,
which
are
needed
for
displaying,
as
a
result,
it
realizes
continuous
and
super
high
waveform
capture
rate.
Average
Times
(when
using
average
sampling)
2
~
512
Set
average
times,
which
steps
from
the
multiple
of
2,
i.e.
2,
4,
8,
16,
32,
64,
128,
256
and
512.
Use
“MULTIPURPOSE”
knob
to
change
average
times.
2.5.1
Acquiring
Mode
Change
the
acquiring
mode
settings
and
observe
the
consequent
change
in
waveform
display.
Normal
Sampling:
Under
this
mode,
oscilloscope
takes
signal
samples
at
an
equal
time
interval
so
as
to
rebuild
a
waveform.
For
most
waveforms,
it
ensures
best
displaying
effect
with
the
use
of
mode.
Peak
Detection
Under
this
mode,
oscilloscope
takes
both
the
maximum
and
minimal
signal
values
within
the
sampling
interval
so
as
to
acquire
signal
envelopes
or
narrow
pulses
that
might
be
lost.
Though
the
use
of
this
mode
can
avoid
signal
aliasing,
the
noise
displayed
is
relatively
large.
Under
this
mode,
oscilloscope
can
display
all
pulses
with
pulse
width
at
least
equals
to
sampling
period.
The
use
of
peak
detection
mode
can
effectively
reduce
the
appearance
of
false
waveform.
Average
Under
this
mode,
oscilloscope
will
average
the
waveform
samples
taken
repeatedly
to
reduce
the
random
42
Содержание UTD2000M
Страница 1: ...UTD2000CM User Manual ...
Страница 77: ...UTD2000M User Manual Fig 3 16 Pass Fail Detection Setting 1 Fig 3 17 Pass Fail Detection Setting 2 76 ...
Страница 92: ...UTD2000M User Manual This user manual may be revised without prior notice 91 ...
Страница 93: ...UTD2000M User Manual 92 ...
Страница 94: ...UTD2000M User Manual 93 ...