UTD2000M
User
Manual
menu
settings
Description
Menu
Settings
Description
Coupling
AC
DC
EARTH
Block
the
DC
component
in
an
input
signal
Allow
the
passage
of
both
AC
and
DC
components
in
an
input
signal
。
Disconnect
input
signal
Probe
Current
Voltage
Probe
Type
Settings
1
×
10
×
100
×
1000
×
User
‐
defined
Based
on
the
attenuation
coefficient
of
the
probe
to
select
a
value
in
order
to
maintain
accurate
reading
for
vertical
deflection
factor.
Totally,
there
are
five
options
to
choose
from,
which
are
1
×、
10
×、
100
×、
1000
×
and
user
‐
defined.
Bandwidth
Limitation
Open
Close
Limit
bandwidth
to
20MHz,
so
as
to
reduce
the
displayed
noise
Full
bandwidth
Phase
Reversed
On
Off
Turn
on
the
reversed
waveform
function;
Normal
display
of
waveforms
Volts/Grids
Rough
Fine
Follow
1
‐
2
‐
5
stepping
to
set
vertical
deflection
factor
during
rough
tuning
Further
division
within
the
scope
of
rough
tuning
can
be
done
in
fine
tuning
so
as
to
improve
vertical
resolution.
2.2.1
Channel
Coupling
Settings
Take
the
application
of
signal
in
CH1
as
an
example,
for
which
the
signal
under
test
is
a
sinusoidal
signal
containing
DC
component.
Press
F1
and
then
use
“MULTIPURPOSE”
knob
to
select
AC
1M
Ω
,
whose
channel
coupling
is
set
as
AC
coupling
mode.
Thus
the
DC
component
contained
in
the
signal
under
test
is
blocked.
Waveform
display
is
shown
as
below.
AC
Coupling
Settings
21