User’s Manual MSO/UPO2000 Seri es
69 / 137
Instruments.uni-trend.com
Waveform when the average mode is not adopted Waveform when the 32 times average mode
Note
: Average and high resolution uses different average methods. The former is multiple sampling average,
the latter is single sampling.
10.3 Storage Depth
The storage depth is the number of waveform that can be stored in the oscilloscope during a trigger
acquisition. It reflects the memory storage capacity of the acquisition.
The relation of storage depth, sampling rate and waveform length should meet the calculating formula:
storage depth = sampling rate×horizontal time base× the number of grid of horizontal direction
MSO/UPO2000 the maximum storage depth 56 Mpts (per channel). In ACQUIRE menu, user can freely to
set the storage depth to auto, 700, 7 k, 70 k, 700 k, 7 M, 14 M, 28 M, 56 M.The default setting is auto.