User’s Manual MSO/UPO2000 Seri es
60 / 137
Instruments.uni-trend.com
8.1 Parameter Measurement
MSO/UPO2000 series oscilloscope can automatically measure 36 kinds of parameters. It includes voltage,
time and other.
①
Voltage
Max
:
The voltage from the highest point of the waveform to GND.
Min
:
The voltage from the lowest point of the waveform to GND.
High
:
The voltage value from the flat top of the waveform to GND.
Low: The voltage value from the bottom of the waveform to GND.
Middle: Half of the sum of the voltage values at the top and bottom of the waveform
Pk-Pk: Peak-to-peak value, the voltage value from the highest point to the lowest point of the waveform.
Amp
:
Voltage from top to bottom of the waveform.
Mean
:
The average amplitude of the waveform in the screen.
CycMean
:
Average amplitude of the waveform in one period.
RMS: The effective value. The energy generated by the conversion of AC signal, it corresponds to the DC
voltage that generates equivalent energy.
CycRMS: The energy produced by the conversion of AC signal in a period, it corresponds to a DC voltage
that generates equivalent energy.
AC RMS: The RMS value is the waveform which DC component has removed.
②
Time