User’s Manual MSO/UPO2000 Seri es
34 / 137
Instruments.uni-trend.com
Figure 6-6
Based on sin
θ
=A/B or C/D,
θ
is the phase angle between channels, the definition of A, B, C, D see Figure 6-6.
So the phase angle is
θ
=±arcsin
(
A/B
)
or
θ
=±arcsin
(
C/D
)
.
If the main spindle of elliptical within I, III quadrant, then the acquired phase angle should at I, IV quadrant,
that is within
(
0
~
π/2
)
or
(
3π/2
~
2π
)
.
If the main spindle of elliptical within II, IV, then the acquired phase angle should within
(
π/2
~
π
)
or
(
π
~
3π/2
)
.
In addition, if the frequency or phase difference of the two signal to be measured are integer times,
calculating the frequency and phase relation of the two signals based on the Figure 6-7.
Figure 6-7
6.5 Multi-Scopes
In multi-scopes, CH1~CH4 can be set to different time base scale, so that signals of different frequency in
multiple channels can be observed at the same time. Enter Multi-Scopes interface via horizontal menu.
Switch Multi-scopes to enter this mode as shown in Figure 6-8. Each channel can be set to different
frequency, amplitude and waveform and all can be stable generated in multi-scopes.
In multi-scopes, it can turn on, off, activate channel. Time base scale, volts/div scale, horizontal
displacement, vertical displacement, trigger setting of channel can be adjusted independently.
Multi-Scopes supports split the screen to display the current channel on the upper part for observing
waveform, as shown in Figure 6-5.