ULVAC CRYOGENICS INCORPORATED

 

close the right panel. 

(5)  Press RUN/STOP button of EMP-14A to reboot the entire system. 

 

(For the system with IM-120

 

(1)  Open the dry air valve to supply dry air to IM-120. 

(2)  Supply nitrogen gas to EMP-14A for at least one hour. 

(3)  Press the RUN/STOP switch of EMP-14A to reboot the entire system. 

 

NOTE: Please inform us in advance when the system is to be suspended for longer 

than two weeks. 

 

8.5  Corrective action in the event of an error 

In an event of an error, follow the procedure below and in “13.Trouble Shooting”. 

 

8.5.1 

Corrective action when electric power failure occurs 

EMP-14A and GN-10i  will  be  automatically  restored  from  electric  power  failure 

and they will be rebooted automatically when the electricity comes back. However, 

when the amount of liquid nitrogen inside is 75% or more, production may not be 

resumed at the time of rebooting.  (This is  the common precaution when 

rebooting). 

 

8.5.2 

Corrective action for COMP. AL. 

(1) Check the electric supply voltage and capacity. 

(2) Remove the acrylic panel of EMP-14A compressor display, turn down the lever 

of compressor OPERATION switch and return it to upward again. 

(3) Press RESET button of EMP-14A. 

(4) When a trouble is temporary, SA115-C re-starts. When the system stops due 

to abnormal temperature, it may unable to resume for about 30 minutes until 

the detection part is cooled down.  When  SA115-C  does not start at all with 

COMP.ALARM  illuminated,  refer to "13.Trouble  Shooting"  and perform 

corrective actions. 

 

8.5.3 

Corrective action for GN2 AL. 

(For the system with GN-10i)

 

(1) Turn off the RUN/STOP switch of EMP-14A. 

(2) Reset the GN-10i. 

(3) Turn on the RUN/STOP switch to re-start EMP-14A. 

(4) When nitrogen gas begins to flow from GN-10i, GN2 AL. lights off. 

Содержание EMP-14A

Страница 1: ...al sensors the cryocooler falls under row 6 A 2 d 2 of the control list established by The Wassenaar Arrangement which is equal to row 10 2 of appended table 1 of Japan s Export Trade Control Order Cu...

Страница 2: ......

Страница 3: ...close at hand so that you can use for reference during operation If you purchased our other products and or optional devices with this product read relevant instruction manuals carefully 1 It is stri...

Страница 4: ...caution describes safety hazards or unsafe practices which could result in personal injury or equipment damage Toxic gas or chemicals used There is a risk of severe injury upon contact Corrosive chemi...

Страница 5: ...of EMP 14A front door and lower panel Figure 2 3 11 2 4 Inside the lower panel on the back 11 2 5 EMP 14A Electrical circuit Figure 2 4 12 2 6 EMP 14A Top side Figure 2 5 12 3 Flow Diagram 13 4 Specif...

Страница 6: ...lly 26 9 1 Connecting liquid nitrogen flexible hose 26 9 2 Dispense liquid nitrogen 26 9 3 Disconnecting flexible hoses 27 10 Automatic Transfer of Liquid Nitrogen 28 10 1 Inspections before starting...

Страница 7: ...he oxygen concentration of 18 or below is defined as an oxygen depleted state Choose a well ventilated and good air circulated location to install the present unit and install a ventilator with a capa...

Страница 8: ...of burns if it is touched directly Carry out maintenance repair etc at least 30 minutes after the unit is stopped In this unit high pressure Helium gas is filled To disassemble or dispose of the unit...

Страница 9: ...s regulations or guidelines when disposing of our products We provide Safety Data Sheet SDS of our products upon your request Please contact us when necessary WARNING When the refrigerator unit may ha...

Страница 10: ...4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...

Страница 11: ...his equipment please refer to the operation manual for SA115 C 1 3 PSA Pressure Swing Adsorption Nitrogen Generator type GN 10i PSA Nitrogen Generator GN 10i generates highly concentrated nitrogen fro...

Страница 12: ...D 2 Component Description Figure 2 1 Front view of EMP 14A Figure 2 2 Operation panel of EMP 14A Emergency Off Button Operation Panel Pressure Gauge LN2 Manual Dispense Port SA115 C Display Air Entran...

Страница 13: ...the port may reach extremely low temperature use cautions not to touch such port during or right after dispensing liquid nitrogen Pressure gauge Displays the pressure inside the dewar SA115 C display...

Страница 14: ...ispense port To be referred to as Manual supply The manual supply is terminated automatically in 10 minutes This button lights green during manual supply and blinks from 30 seconds before the automati...

Страница 15: ...OGENICS INCORPORATED 9 Maintenance This light illuminates when the maintenance interval of GN 10i is approaching Temperature meter Temperature meter is an optional It displays the temperature inside t...

Страница 16: ...and lower panel Figure 2 4 Electrical circuit Figure 2 5 Top of EMP 14A Level sensor converter Fuse Circuit breaker PLC Terminal block Automatic transfer port Pressure gauge Liquid nitrogen dewar Cold...

Страница 17: ...t for charging helium gas Do not touch except when needed as touching it may result in helium leakage OPERATION switch Use this lever to switch between start and stop of SA115 C Normally move the leve...

Страница 18: ...ignal from each sensor In the event of an error please inform the status of input and output lights of the sequencer when contacting us Level sensor converter An electric component for detecting the a...

Страница 19: ...ol Circuit Temperature Indicator TI Optional Pressure Meter Pressure Relief Valve Flexible Hose LN2 Dispense Ambient Temperature Solenoid Valve Pressure Control Valve Coldhead Cable Helium Gas Tube He...

Страница 20: ...han LN2 manual dispense port EMP 14A is for indoor use only As EMP 14A is air cooled make sure that the air intake and exhaust ports are always open not blocked 4 2 PSA Pressure Swing Adsorption nitro...

Страница 21: ...0i Voltage 100VAC 10 single phase Power capacity 15A or more Power consumption Approx 400W 50 60 Hz Connection 3 prong plug 5 3 Dry air for IM 120 when using IM 120 Pressure 0 5MPa 1 0MPa Flow rate 12...

Страница 22: ...lities should be located nearby 2 The ambient temperature should be from 10 C to 35 C and humidity is 80 or below 3 Keep the following spaces for maintenance Spaces other than the back are required on...

Страница 23: ...dry air supply Connect the other side of the tube to AIR IN connector on the back side of IM 120 6 4 Utility connection of EMP 14A Remove the back panel of EMP 14A and locate holes for utility connec...

Страница 24: ...ircuit breaker and supply electricity by turning on the breaker in the electric circuit part of EMP 14A Be sure to confirm that the power is supplied to the sequencers or other parts in the electric c...

Страница 25: ...site side of nitrogen gas tube to the nitrogen gas supply valve on the generator 6 4 3 Connection of Nitrogen gas generator remote cable Connect a remote cable to the remote connector located inside o...

Страница 26: ...A B 0 5SQ P24V 011 0 5SQ 1 2 SV1 SV2 R2 Name Symbol Earth Leakage Breaker Terminal Cover Noise Filter Power Supply Programmable Controller Capacitance Level Meter Converter Level Indicator Temperature...

Страница 27: ...M Display N2 GAS ALARM COMP Startup signal Reserve Cryogenic Manual Electromagnetic Valve Cryogenic Auto Electromagnetic Valve Room Temperature Electromagnetic Valve Cooling Water Electromagnetic Valv...

Страница 28: ...en it starts operation normally the POWER switch on EMP 14A flashes with long term light on and short term light off Note Once operation starts RUN STOP switch does not work for the next six minutes I...

Страница 29: ...it reaches to 100 8 3 Shutdown When EMP 14A is turned OFF with the RUN STOP button the RUN STOP light flashes with shorter illumination and longer off for 10 seconds and all the operations stop includ...

Страница 30: ...when the amount of liquid nitrogen inside is 75 or more production may not be resumed at the time of rebooting This is the common precaution when rebooting 8 5 2 Corrective action for COMP AL 1 Check...

Страница 31: ...0 2MPa and flow rate is 10NL min turn ON the EMP 14A RUN STOP switch to restart 8 6 Emergency off In case of an emergency press the emergency off button on the front panel to disconnect the control po...

Страница 32: ...oduction of liquid nitrogen stops during manual supply operation When nitrogen gas supply is suspended liquid nitrogen cannot be dispensed When the EMP 14A is in standby for automatic transfer when th...

Страница 33: ...re to connecting Push down the outer ring of the EMP 14A side joint and pull out the flexible hose CAUTION When disconnecting flexible hoses make sure that they return to room temperature Note that di...

Страница 34: ...ic transfer will also be terminated when the RUN STOP or AUTO READY of EMP14A are turned OFF the LN2 level of EMP 14A is 0 or supply of nitrogen gas stops Liquid nitrogen cannot be dispensed manually...

Страница 35: ...es Manual dispense is not available during eco mode To dispense manually reset eco mode and wait for 6 minutes until the nitrogen gas generator resumes operation When the liquid nitrogen level is lowe...

Страница 36: ...urs or 2 years Shorter 20 000 hours 3 Coldhead 1 Oil adsorber Helium pipe joint GN 10i Air Compressor GN 10i Filter 2 GN 10i Solenoid valve Parts 1year 2years 5years IM 120 Membrane IM 120 Prefilter I...

Страница 37: ...mer inspection The EMP 14A is a high pressure manufacturing equipment Please follow the applicable laws or regulations when operating and maintaining this equipment Be sure to turn off the main power...

Страница 38: ...op button is pressed Reset the emergency stop switch by turning it to the right and restart 2 Equipment fails to start even if the RUN STOP switch lights up LN2 level is not decreased to the re starti...

Страница 39: ...um gas to the appropriate pressure Contact us for more information on charging 6 RUN STOP switch blinks and EMP does not start LN2 production Alarm light is not lit RUN STOP switch is pressed less tha...

Страница 40: ...the helium gas lines and refrigerator cables properly The level sensor fails to detect correctly Please contact us 9 The rate of Liquid Nitrogen production declines Cold head and oil adsorber exceed...

Страница 41: ...on blinks Only 30 seconds left until the time limit of 10 minutes of manual dispense Press MANU START button again to continue dispending manually 12 AUTO READY button and AUTO SUPPLY button blink alt...

Страница 42: ...Gas Tube 1 3m Single Head Wrench 1 Glass Fuse 3 3A x 2 2A x 1 Instruction Manual 1 This book GN 10i Remote Cable for Nitrogen Generator 1 5m Instruction Manual 1 IM 120 Remote Cable for Nitrogen Gene...

Страница 43: ...nts and on site visits for the following seven conditions In those cases the cost shall be your own expense even though the product is within the warranty period 1 Breakdowns due to improper storage o...

Страница 44: ...ed to your company or your customers by failures of ULVAC CRYOGENICS products and compensation for damages to products other than ULVAC CRYOGENICS products and other services are not covered under war...

Страница 45: ...Chigasaki Kanagawa 253 0085 Japan Sales Tel 81 467 85 8884 Service Engineering Division Tel 81 467 85 9366 Fax 81 467 83 4838 ULVAC CRYOGENICS KOREA INC www ulvac cryo co kr 107 Hyeongoksandan ro Che...

Страница 46: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...

Страница 47: ...2019OR04 First edition The model of PSA nitrogen gas generator has been changed The format for Warnings and Cautions has been changed Disposal Considerations has been added 15 Warranty Company addres...

Страница 48: ...R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...

Отзывы: