8
ULVAC CRYOGENICS INCORPORATED
transfer function. While the AUTO READY is ON, the
EMP is in stand-by mode for automatic transfer, and
automatic transfer to the destination set by receiving
the signal is possible. While standing by, the button
lights green. When the level sensor in the transfer
destination detects any trouble, green light blinks.
AUTO SUPPLY
While using the optional automatic transfer function
with our level sensor, press this button to start
transfer while in stand-by mode. When automatic
transfer is activated, the button lights white.
MANU START
Press this button to manually dispense liquid
nitrogen from LN2 manual dispense port. (To be
referred to as “Manual supply”) The manual supply is
terminated automatically in 10 minutes. This button
lights green during manual supply, and blinks from 30
seconds before the automatic termination. When
you wish to supply more, press this button again
while it is blinking to extend 10 more minutes.
MANU STOP
Press this button to interrupt the manual supply of
liquid nitrogen supply operation.
Indicator
Displays the guideline amount of liquid nitrogen in
the dewar.
eco
Press this button to switch to eco-mode operation.
The button lights blue when in eco-mode. Refer to
"11. eco mode " for more detail.
COMP. AL.
Lights red when the SA115-C encounters an error.
The RUN/STOP button blinks at the same time.
GN2 AL.
The indicator lights red when the nitrogen gas
generator stops due to faults. Then RUN/STOP
switch blinks as well.