C30VR Compressor Installation, Operation, Maintenance and Troubleshooting
7-4
ULVAC CRYOGENICS INCORPORATED
Figure 7-5
Connecting and disconnecting self-sealing coupling with two
spanners
9. Install the replacement adsorber. Remove the two dust caps from the replacement
adsorber. Put the adsorber into the compressor so as to place it at the mounting tab
on the base of the compressor. Secure the adsorber with the mounting bolt
temporally.
10. Connect self-sealing coupling to the adsorber with the two spanners. Tighten the
coupling referring to Figure 7-5. Recommended tightening torque is 20N
・
m.
Connect the coupling quickly to minimize helium leakage.
11. Put the rear panel back on the compressor to align the helium connector of the
adsorber with the hole of the rear panel and then remove the rear panel again. Fix
the adsorber mounting bolt firmly this time.
TURN with 30mm spanner
HOLD with 26mm spanner
(Self-sealing coupling)
Clockwise: Tightening
Counter Clockwise: Loosening
Adsorber
Flexible hose
Figure 7-6
Remove adsorber mounting bolt
Figure 7-7
Slide adsorber outward
Содержание C30MVER
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 16: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 18: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 32: ...C30VR Compressor Instruction Manual 1 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 108: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 110: ...Revision History R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...