C30VR Compressor Instruction Manual
Table 1-3
Power supply rating
Item
C30MVER
C30MVERT
C30VR
C30MVR
C30VRT
C30MVRT
C30PVRT
Rated working voltage
50Hz
190 – 220 VAC
60Hz
200 – 230 VAC
Power phase
3
Stationary power
Consumption
50Hz
4.3kW 5.0kW 5.7kW
60Hz
5.2kW 6.0kW 7.5kW
Power factor
0.85
Stationary electric current
consumption
(*
1
)
50Hz
16A 17A 19A
60Hz
18A 21A 24A
Locked rotor electric current
131A
151A
Customer-supplied rated electric
current (minimum)
(*2)
30A
/ 5kA
40A
/ 5kA
Grounding
D class grounding
(with the ground resistance of 100
Ω
or less)
(*
1
)
During cooling down of the cryopump, until it stabilizes at a low temperature, the electric current value is
about 10% larger than
the stationary electric current.
Longer flexile hoses also increase the electric current. (A set of 20m flexible hose will increase the electric
current by 5% compared to the nominal value.
(*2)
Circuit Breaker Current Rating / Interrupt Rating
Table 1-4
Cooling water requirements
Item Requirements
Water temperature (Cooling water inlet)
5 – 32
℃
Flow rate
5 - 15 L/min
Cooling water supply pressure
Max. 0.7MPaG
Pressure drop (water inlet to water outlet)
0.02 – 0.17MPa (See Fig. 2-3)
Alkalinity (pH)
6.5 – 8.0 pH
Calcium Carbonate
< 75ppm
Refer to “Section 2.4 Cooling water requirements” for more details.
Table 1-5
Installation environment and operation requirements
Item Requirements
C30MVER,
C30MVERT
C30VR, C30MVR
(For trap operation)
C30VR,
C30MVR
C30VRT,
C30MVRT
C30PVRT
Ambient operating temperature
10 – 35
℃
Ambient storage temperature
-10 – 55
℃
Humidity
5 - 90% (non-condensing)
Altitude <
1000m
Vibration
< 0.6G, 10 – 55 Hz
Atmosphere
Free from metal powder, dust,
flammable and corrosive gas
Static helium pressure
at 20
℃
(*1)
1.3±0.04MPaG 1.2±0.04MPaG
1.4±0.04MPaG
Sound pressure level
during operation
< 70 dBA
(*1)
Refer to “Section5.1 Verification of helium static pressure” for details.
1-6
ULVAC CRYOGENICS INCORPORATED
Содержание C30MVER
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 16: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 18: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 32: ...C30VR Compressor Instruction Manual 1 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 108: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 110: ...Revision History R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...