Section
7
Maintenance
Remove the front panel by
unscrewing six screws.
Helium filling fitting
Helium filling valve knob
Helium pressure gauge
(Top view)
Figure 7-8
Locations of front panel screws and components inside the
compressor
(2) Raising Helium Pressure
When the helium pressure gauge reads a value lower than the prescribed standard value,
adding helium gas is required.
When charging helium or performing helium circuit decontamination, equipments
(regulators, charging hoses, adopters, etc.) that can be used at 2.0MPaG or above are
required.
Table 7-2 shows items and devices required for adding helium gas.
Table 7-2
Equipment requirement for adding helium gas
Item Specification
Helium bottle
99.999% pure or above
Pressure regulator
For Helium gas (left-hand screw),
Lower side gauge of 4-6MPaG
Helium charging hose
Terminating in a 1/4 inch female flare fitting
ULVAC CRYOGENICS INCORPORATED
7-7