Section 4 Equipment Installation, Piping, and Wiring
ULVAC CRYOGENICS INCORPORATED
4-13
NOTE:
Clear identification of solder: Listed below are the recommended abbreviations for
identifying solders. Labeling the kind of solder clearly helps you to have proper
maintenance.
Lead eutectic solder
SnPb
or
SP
Lead-free solder (Sn-Ag)
SnAgCu
or
SAC
Lead-free solder (Sn-Cu)
SnCuNi
or
SCN
Connecting remote cable
Connect the cable plug to the compressor receptacle. After connecting the plug to the
“REMOTE RESPONCE” receptacle on the rear panel of the compressor, turn the rotating
part (coupling nut) of the plug to the right. Turn it until it won’t turn any more to connect
the plug properly.
4.7 Wire rooting for compressor and other equipments
The compressor and other equipments must be wired in accordance with applicable local
and national standards where the equipments are installed as well as international
standards (e.g., NEC/CEC). Signal wires, sequence control wires, and power wires for
connecting equipments must be wired in separate conduits to prevent interference during
the operation.
In addition, analog wire is sensitive to and susceptible to noise, it must be separated from
other wirings in the method above.
If it is impossible to wire in separated conduits, keeping enough distance between those
wires can be effective as “separating wires”. Interference between wires depends on
electromagnetic force but in general it is effective to keep the distance of 300mm or more
between each wiring.
Note: Electromagnetic interference where the C30VR system, consists of the compressor
and the refrigerator, effect to other equipments is not critical, however if the wiring of
C30VR system is not separated from other equipments it may cause interference to
sensitive sensor wires, analog wires, and other high impedance wiring of other
equipments.
Connector must be soldered by authorized personnel familiar with the kinds of solder and
soldering procedure. Failure to observe this warning may result in breakdown of the
product, or decrease the life of soldered parts severely.
CAUTION
The remote wires must not be either put in the same conduit or routed with the power wires.
It may cause interference with the compressor operation. Failure to observe this warning
may result in damage to and/or break down of the equipment.
CAUTION
Содержание C30MVER
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 16: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 18: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 32: ...C30VR Compressor Instruction Manual 1 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 108: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 110: ...Revision History R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...