Chapter 3: System Installation
00.053.205, Revision A SonixMDP/SP/OP Service Manual
3-18
To Configure the UPS:
1.
Using the openings provided to access the UPS breakers, turn
1: Battery Breaker
to the ON position.
2.
Turn
2: AC Input Breaker
to the ON position.
3.
Turn
2: AC Input Breaker
to the OFF position.
4.
Turn
2: AC Input Breaker
to the ON position.
5.
Connect the power cord to a wall outlet.
6.
Press and hold the console
POWER
button for one second to turn on the system.
7.
Power off and unplug the system.
Caution:
It is important to turn on the breakers in the order specified.
Note:
Although they are marked "
EMERGENCY SHUTDOWN ONLY
", these openings are also intended to
give breaker access to qualified Ultrasonix Service personnel.
Note:
At this point, depending on the length of time the system remains plugged in and powered on, the fan
may automatically shut down. If this occurs, it does not pose a problem.