SonixMDP/SP/OP Service Manual 00.053.205, Revision A
Chapter 11: Maintenance
11-5
11.2.11 System Filter
This filter should be cleaned approximately every three to six months. Periodically, the system will present a
cleaning reminder message. Always clean the filter when this reminder is presented.
SonixMDP/SP/OP Q+ comes equipped with a double layer filtering system. The first layer, or system filter, is
accessible without removing anything from the system – the filter can simply be pulled out. The second layer (fan
filtering) is attached to the modulo and necessitates modulo removal.
Cleaning the system filter every three to six months helps to safely prolong the modulo fan filter cleaning interval
(refer to
for the various maintenance procedure frequencies).
Use a vacuum cleaner to remove the dust from the system exhaust filters or vents. If a vacuum cleaner is not
available, dust off by hand or with an anti-static brush.
11.2.11.1Cleaning the System Filter
If the system (fan) filter is cleaned early (ie, before the system (fan) filter cleaning reminder is displayed), the
cleaning reminder clock should be reset.
To Reset the System Filter Cleaning Reminder:
1.
Press the console
button.
2.
Select
Administrator > Service
.
3.
Enter the
Level 1
,
Service Mode Password
and select
OK
.
4.
From the
Advanced Tools and Options
drop-down menu, select
Reset Fan Filter Reminder
.
5.
Select
Run > Close
.
Cautions:
Power off and unplug the system prior to cleaning.
Vacuum only. DO NOT apply any liquids to the system filter.
Filter cleaning frequency is dependant upon usage location. If the system is used in a high traffic area (such as an
Emergency Room), the filter may require more frequent cleaning.
Failure to regularly clean the system filter may cause reduced air flow and result in the system overheating.
System failures due to a lack of filter cleaning may not be covered by the Warranty or a Service Contract.
Warning:
Do not perform any internal system maintenance if the UPS breakers are turned on.
Caution:
Power off and unplug the system prior to cleaning the fan filters.
Note:
To obtain a
Level 1
,
Service Mode Password
, contact Ultrasonix Technical Support.
Содержание Sonix MDP
Страница 1: ...Analogic Ultrasound Service Manual SonixMDP SP OPQ Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 205 Revision A SonixMDP SP OP Service Manual 2 2...
Страница 46: ...Chapter 5 Software 00 053 205 Revision A SonixMDP SP OP Service Manual 5 4...
Страница 120: ...Chapter 8 DICOM 00 053 205 Revision A SonixMDP SP OP Service Manual 8 24...
Страница 147: ......