Chapter 3: System Installation
00.053.205, Revision A SonixMDP/SP/OP Service Manual
3-14
Equipment/Tools Required:
• fine tipped, narrow shafted, flat head screwdriver (e.g.,
⅛
" (3mm) jeweler's screwdriver).
To Change Voltage Settings:
1.
Power off and unplug the power cord from both the wall outlet and the system.
2.
Use the screwdriver to open the fuse box lid.
3.
Position the screwdriver at the top (as in
), using it to remove the fuse box.
4.
Flip the fuse box and reinsert it so the correct voltage is visible.
5.
Close the fuse box lid.
Warning:
Do not perform any internal system maintenance if the UPS breakers are turned on.
Содержание Sonix MDP
Страница 1: ...Analogic Ultrasound Service Manual SonixMDP SP OPQ Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 205 Revision A SonixMDP SP OP Service Manual 2 2...
Страница 46: ...Chapter 5 Software 00 053 205 Revision A SonixMDP SP OP Service Manual 5 4...
Страница 120: ...Chapter 8 DICOM 00 053 205 Revision A SonixMDP SP OP Service Manual 8 24...
Страница 147: ......