SonixMDP/SP/OP Service Manual 00.053.205, Revision A
Chapter 4: Performance Testing
4-1
CHAPTER 4: PERFORMANCE TESTING
This section describes the various tests performed on the system immediately following installation, updating and repair. The
purpose of performance testing is to verify the correct operation of both hardware and software as well as mechanical items. During
these tests, the system should be running in normal operating mode.
4.1
CONSOLE TEST
To ensure the console is functioning properly, test the console and touch screen buttons/dials listed in
and verify
the operation corresponds to the expected result. Note and report any operations that do not function correctly.
Figure 4-1: Console
Table 4-1: Console
Note:
Between each test, return the system to
B-Mode
(press the console button).
1
QSONIX
button
7
PW
button/dial
2
SELECT
button
8
B-MODE
button/dial
3
UPDATE
button
9
M-MODE
button
4
FREEZE
button
10
Touch screen
5
TRANSDUCER
button
11 AUDIO
dial
6
MENU
button
12
Touch screen dials. Each dial is associated with the touch
screen buttons that are directly above the dial.
1
2
3
4
7
6
5
8
9
10
12
11
B
Содержание Sonix MDP
Страница 1: ...Analogic Ultrasound Service Manual SonixMDP SP OPQ Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 205 Revision A SonixMDP SP OP Service Manual 2 2...
Страница 46: ...Chapter 5 Software 00 053 205 Revision A SonixMDP SP OP Service Manual 5 4...
Страница 120: ...Chapter 8 DICOM 00 053 205 Revision A SonixMDP SP OP Service Manual 8 24...
Страница 147: ......