SonixMDP/SP/OP Service Manual 00.053.205, Revision A
Chapter 11: Maintenance
11-1
CHAPTER 11: MAINTENANCE
This section is intended to assist in the effective care, cleaning and disinfection of the system. It is also intended to protect the
system and transducers against damage during cleaning/disinfection.
Be sure to carefully follow all recommendations in this section to maintain the highest standard of cleanliness/disinfection, while at
the same time ensuring the operational safety of all equipment.
11.1 RECOMMENDED FREQUENCY OF MAINTENANCE PROCEDURES
The frequency of preventive maintenance performed on the system plays a key role in eliminating or extending the periods
between downtime due to poor performance or unexpected breakdown. The following table offers recommendations that
must be weighed by factors like frequency of use and environmental conditions. In every case, frequent checks of safety-
related items are highly recommended.
Table 11-1: Maintenance Procedure Frequency
11.2 CLEANING SYSTEM COMPONENTS
Ultrasonix recommends the following cleaning instructions for all external surfaces, including the cart, cables and connectors.
Test/Clean Frequency
Interval
Task
Transducers
Six months
Check for cracks or bent pins.
Safety
Six months
Ground impedance/leakage test.
System Filter
Four months
or as required
Check for good air flow without excessive noise.
Remove and vacuum.
Note:
Filter cleaning frequency is dependant upon usage location. If the system is used
in a high traffic area (such as an Emergency Room) filters may require more frequent
cleaning.
Modulo Fan
Six months
or as required
Remove and vacuum.
System Fans
Six months
Check for good air flow without excessive noise.
Cart
Clean as necessary
The wheels have sealed bearings, therefore no lubrication is necessary.
Electronics
Six months
To be performed only by qualified service personnel.
Periodic
Maintenance (PM)
As prescribed by
Ultrasonix procedures
To be performed only by qualified service personnel.
Cautions:
Power off and unplug the system before cleaning.
Do not spill or spray water on the controls, transducer connection receptacle, or transducer ports.
Содержание Sonix MDP
Страница 1: ...Analogic Ultrasound Service Manual SonixMDP SP OPQ Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 205 Revision A SonixMDP SP OP Service Manual 2 2...
Страница 46: ...Chapter 5 Software 00 053 205 Revision A SonixMDP SP OP Service Manual 5 4...
Страница 120: ...Chapter 8 DICOM 00 053 205 Revision A SonixMDP SP OP Service Manual 8 24...
Страница 147: ......