MD40
PRODUCT INSTRUCTION MANUAL
MD40
Page 15 of 56
Author: N.Carapiet | Revision: 1.1
STAGE 2
| Mounting the drive Flange Assembly
How to install the Flange Assembly:
1.
Use an appropriate CONFLAT Gasket
2.
Use 6 appropriate length Mounting Bolts and washers
(note: only hex head bolts can be
used)
3.
Place CONFLAT Gasket into flange counter bore
4.
Insert MD40 flange assembly
5.
Insert bolts and tighten.
Note: when tightening bolts, care must be taken not to tighten them
unevenly as this may cause the gasket not to seal properly.
CONFLAT Gasket
CF40 Chamber Port
Mounting Bolts
Flange Assembly
Содержание MD40 Series
Страница 41: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 41 of 56 Author N Carapiet Revision 1 1 3 1 3 Switch Pin Out...
Страница 43: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 43 of 56 Author N Carapiet Revision 1 1 3 2 3 Home Sensor Pin Out...
Страница 47: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 47 of 56 Author N Carapiet Revision 1 1 Drawings Base Drive...
Страница 48: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 48 of 56 Author N Carapiet Revision 1 1 Dual Shaft...
Страница 49: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 49 of 56 Author N Carapiet Revision 1 1 Pneumatic Drive...
Страница 51: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 51 of 56 Author N Carapiet Revision 1 1 Side Mounted DC Motor Options...
Страница 55: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 55 of 56 Author N Carapiet Revision 1 1 NOTES...