• If the sensitivity is too low, a blinking arrow pointing to the right will appear and there will be
no numeric decibel visible in the display panel. If this occurs, increase the sensitivity until the
arrow disappears (in low level sound environments the arrow will blink continuously and It
will not be possible to achieve a dB indication until a higher intensity level is sensed).
•
If the sensitivity is too high, a blinking arrow pointing to the left will appear and there will be
no numeric decibel visible on the display panel. Reduce the sensitivity until the arrow
disappears and the numeric decibel value is shown.
NOTE:
The blinking arrow indicates the direction in which the Sensitivity Control Dial is to be turned
when out of range.
•
The Sensitivity Control Dial controls the bar graph display
.
Frequency
This instrument is set to the peak frequency response of the transducers which is 40 kHz. It is non-
adjustable.
To store a reading
There are 2 types of storage modes: Normal and Quick.
For “Normal” Storage
• Firmly “Click”(press in) the Sensitivity Dial. The Storage location will blink and the phrase
SPIN/ CLICK will appear on the bottom of the display panel.
• If you wish to use a storage location other than the one shown, “spin” the Sensitivity dial up
(clockwise) or down (counter clockwise) to the desired location.
• If the Storage Location is the one you chose to use, click the Sensitivity Dial again and you
will see a prompt on the bottom of the display panel: STORE? YES . If you want to store the
data, “click” the Sensitivity Dial once more and the record is stored at the set location. The
Storage Location number will automatically move up to the next sequential number.
• If you chose not to store the record, “spin” the Sensitivity Dial and you will see the word NO,
“click” the Sensitivity Dial and you will return to the operation mode.
• For Quick Store (see Set Up Mode, “Menu 05; Store Mode”)
• When in the Quick Store mode, “click (press in) the Sensitivity Dial once and the record is
stored. The Storage Location number will automatically move up to the next sequential
number.
To overwrite data or to enter data in a new location
•
Click (press in) the Sensitivity dial button; the Storage Location number will blink.
• Spin the Sensitivity Dial until the desired storage location is displayed on the screen
• Click the Sensitivity Dial again and the prompt STORE YES? Will appear.
• To store the new information in that location, “click” the Sensitivity Dial again and the record
will be over written.
To download the information
• Refer to Setup Mode, 01 Send Records