Version 1
16
NOTE:
To prevent damage to the module plug, always remove the module BEFORE attaching
and/or removing the Rubber focusing Probe.
Stethoscope module
• The metal rod acts as a wave-guide, directing structure borne ultrasounds directly to the
receiving transducer with little impedance.
• Align the plug located at the rear of the module with the receptacle in the front end of the
Pistol Housing and plug in.
• Touch test area.
As with the scanning module, go from the “gross” to the “fine”. Start a maximum sensitivity on the
Sensitivity Dial and proceed to reduce the sensitivity until a satisfactory sound level is achieved.
Stethoscope extension kit
1. Remove the Stethoscope Module from the Metered Pistol Housing.
2. Unscrew the metal rod in the Stethoscope Module.
3. Look at the thread of the rod you just unscrewed and locate a rod in the kit that has the same
size thread -
this is the “base piece”.
4. Screw the Base Piece into the Stethoscope Module.
5. If all 78cm are to be utilized, locate the middle piece. (This is the rod with a female fitting at
one end) and screw this piece into the base piece.
6.
Screw third “end piece” into middle piece.
7.
If a shorter length is desired, omit step 5 and screw “end piece” into “base piece”.
Long range module
• Plug in to front end.
• Align the plug located at the rear of the module with the receptacle in the front end of the
Pistol Housing and plug in.
• Start to scan the test area.
RAS-MT
The magnetically mounted transducer acts as a wave guide. The cable attaches to the RAM
(Remote Access Module) which is plugged into the pistol grip housing.
•
Make sure RAS-MT cable is attached to the RAM
•
Plug RAM into the front end.
•
Place the magnet transducer on the test point