7. Check and verify that the ambient temperature range is 41ºF (5ºC)-
104ºF(40ºC), it is preferable not to exceed 86 ºF (30ºC).
8. Check and verify that the ambient relative humidity does not exceed
80%
9. The operational altitude shall not be over 13123 ft (4000 meters).
10. Ambient pressure shall not be lower than 8.8 psi (60.5 KPa) (if the
altitude and temperature are kept in the manufacturer's instructed
ranges above, and no forced extreme negative pressure is applied near
the autoclave, then ambient pressure of 8.8 psi (60.5 KPa) or higher is
guaranteed, as it is a function of altitude and temperature).
11. Operate the autoclave only in the manner specified in the manual. If the
equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
5.3 Connections to Utility Supplies
1. Check and verify that the power supply is a 1 phase, 120VAC ±10%,
60Hz, 15A supply.
2. Check and verify that the electrical net is protected by a current leakage
safety relay.
5.4 Storage
After the removal of the autoclave from the package, we recommend the
following:
Keep the device dry.
Keep the device away from sunlight and protect it from heat.
Page 22
Содержание AMS11-120-T
Страница 2: ......
Страница 4: ...https www tuttnauerusa com...
Страница 16: ...1 10 Device Specifications 1 10 1 Overall Dimensions Page 9...
Страница 33: ...2 Select the Version information 3 The Version information is displayed For an example see below Page 26...
Страница 56: ...Note For the results of the cycle see description in section 8 5 Page 49...
Страница 67: ...Press to receive the entire printout of the selected cycle history Page 60...
Страница 86: ...Page 79...
Страница 105: ...Cat No MAN205 0502003EN Rev C COPYRIGHT...