9.4.3
Replacing the Drain Filter
Note:If after successive cleaning of the filter the result is not satisfactory
proceed to the replacement of the drain filter (paragraph 9.4.3 below).
Cautions!
Before proceeding, make sure that the electric cord is
disconnected and there is no pressure in the chamber.
Use scissors to open the filter bag and not sharp blades
or pointed instrument.
Check that the new filter has not exceeded the maximum
shelf life.
Carefully un-pack the new filter and examine it for any
signs of damage.
Remove any protective packaging before inserting the
filter into place.
Carefully Insert the new filter into the housing. Do not
force.
Caution!
Make sure the autoclave is not hot before cleaning it
The DRAIN filter (1) is located inside the autoclave chamber at the bottom far
end, to reach the filter open the autoclave chamber door, remove the tray (see
below).
1. Replace the old DRAIN filter, by the new DRAIN filter (P/N: CMT511-0048).
2. Open the autoclave chamber door, remove the tray (2), (see below).
3. Unscrew the old Drain filter assembly, using M14 wrench.
Page 68
Содержание AMS11-120-T
Страница 2: ......
Страница 4: ...https www tuttnauerusa com...
Страница 16: ...1 10 Device Specifications 1 10 1 Overall Dimensions Page 9...
Страница 33: ...2 Select the Version information 3 The Version information is displayed For an example see below Page 26...
Страница 56: ...Note For the results of the cycle see description in section 8 5 Page 49...
Страница 67: ...Press to receive the entire printout of the selected cycle history Page 60...
Страница 86: ...Page 79...
Страница 105: ...Cat No MAN205 0502003EN Rev C COPYRIGHT...