10.2 Error messages & Troubleshooting
Message
Description
“Door is Open”
In STBY stage, when Door switch (DI 1)
Open AND Door lock DO 2 is ON
"Analog Input Error"
This message is displayed when any
Temperature sensor or Pressure sensor
is disconnected or out of range.
* The name of the input should appear in
the message
"Chamber temperature not
in range"
This message is displayed if the
temperature in the chamber > End
temperature (248°F) or < 35.6°F
"Chamber pressure not in
range"
This message is displayed if Chamber
pressure <ATM pressure-0.4 bar or
>ATM Pr0.1 bar).
Cancel
"I/O card is not connected"
This message is displayed if I/O card is
disconnected (both while cycle is running
or not).
"Low Temp"
This message is displayed if the
temperature drops for more than 1
second below the sterilization
temperature during sterilization cycle.
"High Temp"
This message is displayed if the
temperature raises 37.4°F above
sterilization temperature during the
sterilization stage for 2 seconds during
sterilization cycle.
"High Temp. (Ending)"
This message is displayed if the system
cannot reach the required temperature, in
the chamber, within 10 minutes.
"Heat Time Error"
This message is displayed if the system
Page 72
Содержание AMS11-120-T
Страница 2: ......
Страница 4: ...https www tuttnauerusa com...
Страница 16: ...1 10 Device Specifications 1 10 1 Overall Dimensions Page 9...
Страница 33: ...2 Select the Version information 3 The Version information is displayed For an example see below Page 26...
Страница 56: ...Note For the results of the cycle see description in section 8 5 Page 49...
Страница 67: ...Press to receive the entire printout of the selected cycle history Page 60...
Страница 86: ...Page 79...
Страница 105: ...Cat No MAN205 0502003EN Rev C COPYRIGHT...